CLIA Kit for Very Low Density Lipoprotein Receptor (VLDLR) Rattus norvegicus (Rat) Sandwich CLIA

VLDL-R

Add to Cart Distributors
Overview
Properties
Share your citation Upload your experimental result Review Leave a message
Loading...

Share a new citation as an author

Upload your experimental result

Review

Please attach serial No. on instruction manual

Contact us

Please fill in the blank.

Name*
Organization
Address
E-mail address*
Telephone
Inquiry*
Verification code* CheckCode
  • CLIA Kit for Very Low Density Lipoprotein Receptor (VLDLR) Packages (Simulation)
  • CLIA Kit for Very Low Density Lipoprotein Receptor (VLDLR) Packages (Simulation)
  • CLIA Kit for Very Low Density Lipoprotein Receptor (VLDLR) Results demonstration
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Very Low Density Lipoprotein Receptor (VLDLR) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Very Low Density Lipoprotein Receptor (VLDLR) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Substrate A 1×10mL Substrate B 1×2mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

CLIA Kit for Very Low Density Lipoprotein Receptor (VLDLR)

Test principle

The microplate provided in this kit has been pre-coated with an antibody specific to Very Low Density Lipoprotein Receptor (VLDLR). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Very Low Density Lipoprotein Receptor (VLDLR). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Very Low Density Lipoprotein Receptor (VLDLR) level in the sample or standard.;