CLIA Kit for Trefoil Factor 3 (TFF3)
HITF; ITF3; TFI; HP1.B; Trefoil Factor 3, Intestinal
- UOM
- FOB US$ 588.00 US$ 840.00 US$ 3,780.00 US$ 7,140.00 US$ 58,800.00
- Quantity
Overview
Properties
- Product No.SCB656Hu
- Organism SpeciesHomo sapiens (Human) Same name, Different species.
- ApplicationsChemiluminescent immunoassay for Antigen Detection.
Research use only - DownloadInstruction Manual
- CategoryTumor immunityGastroenterology
Sign into your account
Share a new citation as an author
Upload your experimental result
Review

Contact us
Please fill in the blank.
Recovery
Matrices listed below were spiked with certain level of recombinant Trefoil Factor 3 (TFF3) and the recovery rates were calculated by comparing the measured value to the expected amount of Trefoil Factor 3 (TFF3) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 83-93 | 90 |
EDTA plasma(n=5) | 90-102 | 95 |
heparin plasma(n=5) | 81-94 | 88 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Trefoil Factor 3 (TFF3) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Trefoil Factor 3 (TFF3) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Trefoil Factor 3 (TFF3) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 89-99% | 82-101% | 87-94% | 79-94% |
EDTA plasma(n=5) | 89-103% | 97-105% | 78-98% | 84-103% |
heparin plasma(n=5) | 93-101% | 88-102% | 95-104% | 88-97% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
Reagents | Quantity | Reagents | Quantity |
Pre-coated, ready to use 96-well strip plate | 1 | Plate sealer for 96 wells | 4 |
Standard | 2 | Standard Diluent | 1×20mL |
Detection Reagent A | 1×120µL | Assay Diluent A | 1×12mL |
Detection Reagent B | 1×120µL | Assay Diluent B | 1×12mL |
Substrate A | 1×10mL | Substrate B | 1×2mL |
Wash Buffer (30 × concentrate) | 1×20mL | Instruction manual | 1 |
Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle
The microplate provided in this kit has been pre-coated with an antibody specific to Trefoil Factor 3 (TFF3). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Trefoil Factor 3 (TFF3). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Trefoil Factor 3 (TFF3) level in the sample or standard.;
Giveaways
Increment services
-
Single-component Reagents of Assay Kit
-
Lysis Buffer Specific for ELISA / CLIA
-
Quality Control of Kit
-
CLIA Kit Customized Service
-
Disease Model Customized Service
-
Serums Customized Service
-
TGFB1 Activation Reagent
-
Real Time PCR Experimental Service
-
Streptavidin
-
Fast blue Protein Stain solution
-
Single-component Reagents of FLIA Kit
-
Streptavidin-Agarose Beads
Citations
- Serum trefoil factor 3 is a promising non-invasive biomarker for gastric cancer screening: A monocentric cohort study in ChinaBiomedcentral: Source
- Serum Trefoil Factor 3 as a Protein Biomarker for the Diagnosis of Colorectal Cancer.pubmed:27760866
- Obstructive sleep apnea and rhonchopathy are associated with downregulation of trefoil factor family peptide 3 (TFF3)-Implications of changes in oral mucus composition.pubmed:29028798
- Protective effects of sinomenine against LPS-induced inflammation in piglets.pubmed:28757275
- Nucleotide-mediated SPDEF modulates TFF3-mediated wound healing and intestinal barrier function during the weaning processPubmed:29555969
- Kidney-based in vivo model for drug-induced nephrotoxicity testingPubmed: 32796873
- Human Synovia Contains Trefoil Factor Family (TFF) Peptides 1–3 Although Synovial Membrane Only Produces TFF3: Implications in Osteoarthritis and Rheumatoid …
- Untersuchungen der Trefoil factor family Peptide 1-3 in humaner Synovialfl¨¹ssigkeit und Synovialmembran und deren Bedeutung bei Arthrose und?¡