Multiplex Assay Kit for Adropin (AD) ,etc. by CBA (Cytometric Bead Array)
ENHO; UNQ470; Energy Homeostasis Associated Protein
- UOM
- FOB US$ 480.00 US$ 499.00 US$ 527.00 US$ 564.00 US$ 601.00 US$ 656.00 US$ 739.00 US$ 924.00
- Quantity
Overview
Properties
- Product No.CMN251Hu
- Organism SpeciesHomo sapiens (Human) Same name, Different species.
- ApplicationsFLIA Kit for Antigen Detection.
Research use only - DownloadInstruction Manual
- CategorySignal transduction
Sign into your account
Share a new citation as an author
Upload your experimental result
Review

Contact us
Please fill in the blank.
Recovery
Matrices listed below were spiked with certain level of recombinant Adropin (AD) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Adropin (AD) ,etc. by CBA (Cytometric Bead Array) in samples.
| Matrix | Recovery range (%) | Average(%) |
| serum(n=5) | 80-94 | 81 |
| EDTA plasma(n=5) | 87-95 | 92 |
| heparin plasma(n=5) | 81-104 | 101 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Adropin (AD) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Adropin (AD) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Adropin (AD) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Sample | 1:2 | 1:4 | 1:8 | 1:16 |
| serum(n=5) | 79-95% | 88-98% | 93-103% | 90-97% |
| EDTA plasma(n=5) | 97-104% | 84-102% | 80-90% | 96-104% |
| heparin plasma(n=5) | 78-94% | 79-101% | 88-95% | 98-105% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
| Reagents | Quantity | Reagents | Quantity |
Assay procedure summary
1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.
Test principle
Anti-Adropin (AD) ,etc. by CBA (Cytometric Bead Array) antibody was coated on the surface of magnetic beads, to make solid phase carrier. Standard or sample were added to the magnetic beads, respectively, in which Adropin (AD) ,etc. by CBA (Cytometric Bead Array) was bound to the antibody connected to the solid phase carrier, and then biotinylated anti-Adropin (AD) ,etc. by CBA (Cytometric Bead Array) antibodies were added. PE-labeled avidin was added. There is a positive correlation between the concentration of Adropin (AD) ,etc. by CBA (Cytometric Bead Array) in the sample and corresponding value of MFI (Median Fluorescence Intensity). Flow Cytometer was used to determine and calculate the sample concentration.
Giveaways
Increment services
Citations
- Alteration of serum adropin level in preeclampsia.pubmed:28501281
- Adropin Is a Key Mediator of Hypoxia Induced Anti-Dipsogenic Effects via TRPV4-CamKK-AMPK Signaling in the Circumventricular Organs of Rats.pubmed:28473751
- Afamin and Adropin in Patients with Alcohol-Induced Liver Cirrhosismanuscript:201803.0258
- Effects of Probiotic Yogurt on Serum Omentin-1, Adropin, and Nesfatin-1 Concentrations in Overweight and Obese Participants Under Low-Calorie DietPubmed: 30232744
- Mitra Zarrati, Mahsa Raji Lahiji, Eisa Salehi, Bahareh Yazdani, Elham Razmpoosh, Raheleh Shokouhi Shoormasti & Farzad ShidfarPubmed: 30232744
- Adropin and apelin‐12 efficiently predict metabolic syndrome in obese childrenPubmed: 32749012
- Effect of Peri-Implant Disease on Adropin Levels: A Cross-Sectional Pilot Study
- Effects of eight weeks exercise training on serum levels of adropin in male volleyball players33794077
