Multiplex Assay Kit for Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array)
HARP; ARP2; Angiopoietin-Related Protein 2
Overview
Properties
- Product No.CMB919Hu
- Organism SpeciesHomo sapiens (Human) Same name, Different species.
- ApplicationsFLIA Kit for Antigen Detection.
Research use only - DownloadInstruction Manual
- CategoryTumor immunityInfection immunityCardiovascular biology
Sign into your account
Share a new citation as an author
Upload your experimental result
Review

Contact us
Please fill in the blank.
Recovery
Matrices listed below were spiked with certain level of recombinant Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) in samples.
| Matrix | Recovery range (%) | Average(%) |
| serum(n=5) | 96-104 | 101 |
| EDTA plasma(n=5) | 85-92 | 89 |
| heparin plasma(n=5) | 78-92 | 87 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Sample | 1:2 | 1:4 | 1:8 | 1:16 |
| serum(n=5) | 80-99% | 83-96% | 93-101% | 88-99% |
| EDTA plasma(n=5) | 93-101% | 81-91% | 84-93% | 90-98% |
| heparin plasma(n=5) | 97-105% | 79-91% | 88-96% | 87-95% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
| Reagents | Quantity | Reagents | Quantity |
Assay procedure summary
1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.
Test principle
Anti-Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) antibody was coated on the surface of magnetic beads, to make solid phase carrier. Standard or sample were added to the magnetic beads, respectively, in which Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) was bound to the antibody connected to the solid phase carrier, and then biotinylated anti-Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) antibodies were added. PE-labeled avidin was added. There is a positive correlation between the concentration of Angiopoietin Like Protein 2 (ANGPTL2) ,etc. by CBA (Cytometric Bead Array) in the sample and corresponding value of MFI (Median Fluorescence Intensity). Flow Cytometer was used to determine and calculate the sample concentration.
Giveaways
Increment services
Citations
- Tumor necrosis factor-α increases angiopoietin-like protein 2 gene expression by activating Foxo1 in 3T3-L1 adipocytesPubMed: 21501655
- Prolyl-4-hydroxylase domain 3 (PHD3) is a critical terminator for cell survival of macrophages under stress conditionsPubmed:24626957
- Lack of Angiopoietin‐Like‐2 Expression Limits the Metabolic Stress Induced by a High‐Fat Diet and Maintains Endothelial Function in MicePubmed:25128474
- Elevated Serum Angiopoietin-like Protein 2 Correlates with the Metastatic Properties of Colorectal Cancer: A Serum Biomarker for Early Diagnosis and RecurrencePubmed:25294915
- Serum angiopoietin-like protein 2 as a potential biomarker for diagnosis, early recurrence and prognosis in gastric cancer patientsPubMed: 26420253
- ANGPTL2 is associated with an increased risk of cardiovascular events and death in diabetic patients.pubmed:27491833
- c-kit Positive Cardiac Outgrowth Cells Demonstrate Better Ability for Cardiac Recovery Against Ischemic Myopathy.pubmed:29238626
- Influence of micro-and macro-vascular disease and Tumor Necrosis Factor Receptor 1 on the level of lower-extremity amputation in patients with type 2 …Pubmed:29879997
- Circulating angiopoietin‑like protein 2 levels are associated with decreased renal function in HIV+ subjects on cART: A potential marker of kidney disease
- Serum tenascin-C is independently associated with increased major adverse cardiovascular events and death in individuals with type 2 diabetes: a French …Pubmed: 32040670